User Manual LW120
SWEep:TIME?
Response and default
The LW120 will return the present sweep time value. The returned value will be in standard scientific format
(for example: 10ms would be returned as 10E-3 – positive numbers are unsigned). Default value is 1e-3.
SWEep:DIRection{UP|DOWN}
Purpose
This command will select the sweep direction up and down.
•
<UP> selects an up direction. The sample clock will sweep from start frequency set by the sclk
parameter to stop frequency set by the stop sclk parameter
<DOWN> selects the down direction. The sample clock will s
•
weep from stop frequency set by the stop
sclk parameter to start frequency set by the sclk parameter
Parameter type
Discrete
SWEep:DIRection?
Response and default
,
The LW120 will return UP or DOWN depending on the present LW120 setting. Default value is UP.
SWEep:SPACing{LINear|LOGarithmic}
Purpose
This command will select the sweep spacing from linear spacing and logarithmic spacing.
•
<LINear> will select linear steps
•
mic steps
<LOGarithmic> will select logarith
Parameter type
Discrete
SWEep:SPACing?
Response and default
N
The LW120 will return LI , or LOG depending on the present LW120 setting. Default value is LIN.
SWEep:TRIGger:MODE(CONTinuous|TRIGered|GATEd}
Purpose
This command will select one of the sweep modes.
•
CONTinuous will select continuous sweep modulation
•
TRIGered will select triggered sweep modulation
•
GATEd will select gated sweep modulation
Parameter type
Discrete
SWEep:TRIGger:MODE?
Remote Programming Reference 4-27