ECO Gold
Access to the main menu level is obtained by pressing the enter key
.
Selection and confirmation of
Setup
Calibration
.
The adjacent menu window appears.
Select
Calibration
with or and confirm with .
The entry window appears. The value indicated on the reference ther-
mometer must be entered as the value.
Change the value with or .
Single figures can be selected by pressing or .
By pressing
(+/-) the arithmetic sign can be changed.
You confirm the set value by pressing .
-
By pressing (ESC) you are returned to the menu level without any change.
13.8
Restoring the factory setting of the internal temperature sensor
If the offset has been adjusted, the factory setting can be restored again.
Access to the main menu level is obtained by pressing the enter key
.
Selection and confirmation of
Setup
Calibration
.
The adjacent menu window appears.
Select and confirm
Factory calibration
with or and .
The adjacent menu window appears.
Select "yes" using or and confirm with to restore the
factory settings.
-
By pressing "no"
(ESC) or you are returned to the menu level without any change.
Summary of Contents for Viscotemp 15 G
Page 77: ...11 Declaration of Conformity 09 2021 ECO Gold 77 134 ...
Page 78: ...78 134 ECO Gold 09 2021 ...
Page 81: ...09 2021 ECO Gold 81 134 W Warnings 53 92 X Xp 106 Xpf 108 ...
Page 82: ...82 134 ECO Gold 09 2021 ...
Page 83: ...09 2021 ECO Gold 83 134 ...
Page 84: ...84 134 ECO Gold 09 2021 ...
Page 85: ...09 2021 ECO Gold 85 134 Appendix with settings ...
Page 121: ...09 2021 ECO Gold 121 134 Click on the tab with the mouse and then on the Device manager ...
Page 129: ...09 2021 ECO Gold 129 134 ...
Page 130: ...130 134 ECO Gold 09 2021 ...
Page 131: ......