60 / 134
ECO Gold
Decalcifying the cooling water circuit
At regular intervals of 3 months or longer (depending on the water hardness / degree of contamination of the cooling
water), the water-cooled condenser must be decalcified or cleaned.
Required equipment:
Two containers of 10 to 20 liters.
Use a suitable pump (drum pump) or possibly use hose with a funnel with funnel located above the cool-
ing water inlet.
Hose between container, pump and cooling water inlet and also between cooling water outlet and container.
Standard
Alternative
Via the water inlet hose, fill the device with decalcifier (pump or hose). Set the set value to 10 °C; after the chiller
starts the water circuit can be filled. Circulate the decalcifier with the pump or continue to top up the decalcifier.
Allow the decalcifier to have an effect (see table below). Then drain the device. Again connect the device to the water
supply and thoroughly flush it out (see table below).
Acting time
Continue with the pumping process until the foaming reaction decays. Generally, this is achieved
after about 20 to 30 minutes.
Decalcifier
LAUDA article number: LZB 126 (5 kg)
When handling the chemicals, the safety information and the instructions for use on the package
are to be followed.
Flushing
Allow at least 10 liters of water to flow through.
Interchange containers
when top one is empty
Cooling water connections
Summary of Contents for Viscotemp 15 G
Page 77: ...11 Declaration of Conformity 09 2021 ECO Gold 77 134 ...
Page 78: ...78 134 ECO Gold 09 2021 ...
Page 81: ...09 2021 ECO Gold 81 134 W Warnings 53 92 X Xp 106 Xpf 108 ...
Page 82: ...82 134 ECO Gold 09 2021 ...
Page 83: ...09 2021 ECO Gold 83 134 ...
Page 84: ...84 134 ECO Gold 09 2021 ...
Page 85: ...09 2021 ECO Gold 85 134 Appendix with settings ...
Page 121: ...09 2021 ECO Gold 121 134 Click on the tab with the mouse and then on the Device manager ...
Page 129: ...09 2021 ECO Gold 129 134 ...
Page 130: ...130 134 ECO Gold 09 2021 ...
Page 131: ......