ECO Gold
125 / 134
Command
Meaning
VERSION_M_3
Query of the software version number of the solenoid valve (shut-off valve 1).
VERSION_M_4
Query of the software version number of the solenoid valve (shut-off valve 2).
VERSION_M_5
Query of the software version number of the high temperature cooler.
VERSION_E
Query of the software version number of the external Pt100 module.
STATUS
Query of the device status 0 = OK, -1 = Error.
STAT
Query of error diagnosis response:
XXXXXXX
X = 0 no error, X = 1 error
1st character = Error
2nd character = Alarm
3rd character = Warning
4th character = Overtemperature
5th character = Low Level
6th character = 0
7th character = External control value missing
RMP_IN_00_XXX
Query of a program segment XXX
(Response: e.g. 030.00_00010_005.00_001.00 => Set-point temperature = 30.00 °C, Time =
10 min, Tolerance = 5,00 K, Pump stage = 1).
RMP_IN_01
Query of the current segment number.
RMP_IN_02
Query of the set program loops.
RMP_IN_03
Query of the current program loops.
RMP_IN_04
Query of to which program further commands refer.
RMP_IN_05
Query of which program is currently running (0 = none).
LOG_IN_00_XXXX
Query of a measurement point XXX from data logger
(Response: e.g. 020.00_021.23_030.50 => Set-point temperature = 20.00
℃
, bath tempera-
ture = 21.23 °C, external temperature = 30.5 °C).
LOG_IN_01
Query of all measurement points from data logger
In contrast to the command "LOG_IN_00" a tabulator character is used here as delimiter instead of
'_'. The measurement points are separated by CR and LF. The end is signaled by CR LF CR LF.
LOG_IN_02
Query of starting time of data logger
(Response: e.g. 20_14_12_20 => Day 20, 14:12:20 hrs.).
LOG_IN_03
Query of acquisition interval from data logger (Response in seconds).
Note:
For ”_“ ” ” (space character) is also admissible.
Unless otherwise stated, with the command the response is always in the fixed-point format "XXX.XX" or for
negative values "-XXX.XX" or "ERR_X" (RS 485 interface, e.g. "A015_XXX.XX" or "A015_ERR_X").
The command from the computer must be terminated with a CR, CRLF or LFCR.
The response from the thermostat is always terminated with a CRLF.
After each command sent to the thermostat, it is necessary to wait for the reply before sending another
command. This ensures that the sequencing of inquiries and answers is clear.
CR = Carriage Return (Hex: 0D); LF = Line Feed (Hex: 0A)
Summary of Contents for Viscotemp 15 G
Page 77: ...11 Declaration of Conformity 09 2021 ECO Gold 77 134 ...
Page 78: ...78 134 ECO Gold 09 2021 ...
Page 81: ...09 2021 ECO Gold 81 134 W Warnings 53 92 X Xp 106 Xpf 108 ...
Page 82: ...82 134 ECO Gold 09 2021 ...
Page 83: ...09 2021 ECO Gold 83 134 ...
Page 84: ...84 134 ECO Gold 09 2021 ...
Page 85: ...09 2021 ECO Gold 85 134 Appendix with settings ...
Page 121: ...09 2021 ECO Gold 121 134 Click on the tab with the mouse and then on the Device manager ...
Page 129: ...09 2021 ECO Gold 129 134 ...
Page 130: ...130 134 ECO Gold 09 2021 ...
Page 131: ......