7.4.2
Setting the temperature set-point value
-
Access to the main menu level is obtained by pressing the en-
ter key .
-
Select the menu point "Set-point temperature" using the en-
ter key .
The entry window appears. The cursor below the temperature value
flashes and can be changed within the displayed limits.
-
Change the value with or .
-
Single figures can be selected by pressing or .
-
By pressing
(+/-), with appropriate equipment, the arith-
metic sign can be changed.
-
Confirm your choice with the enter key .
-
By pressing
(ESC) you are returned to the menu level without any change.
7.4.3
Setting the pump level
With the ECO Vario pump you have six pump levels available with which you can optimize the bath circulation, flow
rate and pressure, the noise generated and the mechanical heat input. With small thermostats (e.g. E 4 G, RE 415 G,
RE 420 G) without an external consumer power levels 1 to 3 are practicable and sufficient.
-
Access to the main menu level is obtained by pressing the en-
ter key .
-
The adjacent menu window appears by selecting and confirm-
ing
Setup
Pump level
.
-
The level can be selected with or . The selected level is
immediately active without confirmation.
-
You quit the menu by pressing
(ESC) or .
Summary of Contents for Viscotemp 15 G
Page 77: ...11 Declaration of Conformity 09 2021 ECO Gold 77 134 ...
Page 78: ...78 134 ECO Gold 09 2021 ...
Page 81: ...09 2021 ECO Gold 81 134 W Warnings 53 92 X Xp 106 Xpf 108 ...
Page 82: ...82 134 ECO Gold 09 2021 ...
Page 83: ...09 2021 ECO Gold 83 134 ...
Page 84: ...84 134 ECO Gold 09 2021 ...
Page 85: ...09 2021 ECO Gold 85 134 Appendix with settings ...
Page 121: ...09 2021 ECO Gold 121 134 Click on the tab with the mouse and then on the Device manager ...
Page 129: ...09 2021 ECO Gold 129 134 ...
Page 130: ...130 134 ECO Gold 09 2021 ...
Page 131: ......