ECO Gold
39 / 134
6.5
Cooling of heating thermostats
At bath temperatures slightly above the room temperature (approx. 2 – 5 K) operation is possible at a low pump level
(1 or 2) without cooling. For temperatures below room temperature cooling must be used.
With the immersion thermostat use a cooling coil (
With bath and circulation thermostats the cooling coil is already built in as standard.
Connect external cooling fluid to the cooling coil. At temperatures above 20 °C, fresh water can be used. Ensure the
lowest possible water consumption.
6.6
First switch-on
Make sure that the details on the name-plate match mains voltage and frequency.
Notice
Use of inadmissible mains voltage or frequency
Property damage
Compare the rating label with the available mains voltage and
frequency.
Note for electric installation on site:
The devices must be protected with a 16 ampere circuit breaker fitted during installation.
Exception: Devices with 13 ampere UK plugs.
Note:
The device mains plug is used as a mains disconnection component.
The mains plug must be easily recognizable and easily accessible.
Only connect units to sockets having a safety earth conductor (PE). No liability is accepted for incorrect mains
connection.
Make sure that if not using an external consumer, the pressure nozzle is closed off or short-circuited to the return
nozzle.
Make sure that the unit is filled according to section (
Summary of Contents for Viscotemp 15 G
Page 77: ...11 Declaration of Conformity 09 2021 ECO Gold 77 134 ...
Page 78: ...78 134 ECO Gold 09 2021 ...
Page 81: ...09 2021 ECO Gold 81 134 W Warnings 53 92 X Xp 106 Xpf 108 ...
Page 82: ...82 134 ECO Gold 09 2021 ...
Page 83: ...09 2021 ECO Gold 83 134 ...
Page 84: ...84 134 ECO Gold 09 2021 ...
Page 85: ...09 2021 ECO Gold 85 134 Appendix with settings ...
Page 121: ...09 2021 ECO Gold 121 134 Click on the tab with the mouse and then on the Device manager ...
Page 129: ...09 2021 ECO Gold 129 134 ...
Page 130: ...130 134 ECO Gold 09 2021 ...
Page 131: ......