Connection tube LZM 045
Screw for adjustment (3x)
Side view
Top view
Starting up of the calibration thermostat
Firstly, assemble the connecting tube LZM 045 as per diagram.
Attention! Do not bend the tube! Hold up with open-end wrench SW 14 mm.
Please move the connecting screws for adjustment of the cylindrical chamber
always gradually (about 2 rotations). Adjust the working chamber as per diagram,
and fix it that way it rises 3 mm above the bath opening. Refill the bath liquid
only up to 15 mm underneath the bath opening and only at working tempera-
ture and switched-off pump.
Recommendation for the adjustment of
the cylindrical working chamber
Level
pump off
Level
pump on