ECO Gold
31 / 134
-
Turn the pump output downwards for external bath circulation.
-
Fit the hose section of the pump connection set onto the outflow elbow
and place the pump connections in the position of the removed blind
flange.
-
Push the holed flange under the pump connections and fasten it with
two cross-head screws to the underside of the control head.
Holed flange
-
Use the flat seal. Make sure the seal is in the correct position. On one
side of the seal there are two steps:
They must be positioned on the side of the display.
-
Refit the control head onto the bath bridge with the two cross-head
screws.
-
Select the division of the pump flow to suit the thermostating task using
the selector switch on the front of the control head.
The position
EXT
gives the greatest flow in the external circuit.
With the position
INT
the external flow is throttled to a minimum and the outlet
for the internal bath circulation is fully opened.
With positions between
INT
and
EXT
the flow is divided up between internal
and external circulation.
Summary of Contents for Viscotemp 15 G
Page 77: ...11 Declaration of Conformity 09 2021 ECO Gold 77 134 ...
Page 78: ...78 134 ECO Gold 09 2021 ...
Page 81: ...09 2021 ECO Gold 81 134 W Warnings 53 92 X Xp 106 Xpf 108 ...
Page 82: ...82 134 ECO Gold 09 2021 ...
Page 83: ...09 2021 ECO Gold 83 134 ...
Page 84: ...84 134 ECO Gold 09 2021 ...
Page 85: ...09 2021 ECO Gold 85 134 Appendix with settings ...
Page 121: ...09 2021 ECO Gold 121 134 Click on the tab with the mouse and then on the Device manager ...
Page 129: ...09 2021 ECO Gold 129 134 ...
Page 130: ...130 134 ECO Gold 09 2021 ...
Page 131: ......