54 / 134
ECO Gold
When the bath temperature is located above the overtemperature switch-
off point, a two-tone alarm sounds. "Overtemperature" appears in the dis-
play, the heater switches off on all poles and the pump and chiller are
switched off via the electronics.
-
Rectify the cause of the malfunction.
-
Wait until the bath temperature has cooled below the over-
temperature switch-off point or set the overtemperature switch-
off point higher than the bath temperature.
If "Overtemperature" appears in the display:
-
Unlock the "Overtemperature" display with .
Before a longer unsupervised operation check the overtemperature protection:
-
Slowly reduce T
max
as described in (
7.4.5). The thermostat should switch off when the actual temperature is
greater than T
max
.
Alarm signaling (see above) must occur.
-
Reset the switch-off point to be higher than the bath temperature.
-
Unlock the "Overtemperature" display with .
8.1.2
Low level: Alarm and checking
When the liquid level falls so far that the heaters are no longer completely
covered with liquid, a two-tone alarm sounds. "Low Level Pump" appears in
the display, the heater switches off on all poles and the pump and chiller are
switched off via the electronics.
-
Rectify the cause of the malfunction.
-
Top up the missing heat transfer liquid (
-
Unlock the "Low Level Pump" display with .
Check the safety system at regular intervals (
8.3.2) by lowering the bath level. Do not carry out this test at a bath
temperature below 0 °C or above 50 °C in order to avoid dangers due to temperatures that are too hot or too cold.
Alarm signaling (see above) must occur.
-
Top up with heat transfer liquid.
-
Unlock the "Low Level Pump" display with .
Switch the device off immediately and withdraw the mains plug if irregularities occur when checking the safety devices.
Contact LAUDA Service (
8.7) or your local service.
Summary of Contents for Viscotemp 15 G
Page 77: ...11 Declaration of Conformity 09 2021 ECO Gold 77 134 ...
Page 78: ...78 134 ECO Gold 09 2021 ...
Page 81: ...09 2021 ECO Gold 81 134 W Warnings 53 92 X Xp 106 Xpf 108 ...
Page 82: ...82 134 ECO Gold 09 2021 ...
Page 83: ...09 2021 ECO Gold 83 134 ...
Page 84: ...84 134 ECO Gold 09 2021 ...
Page 85: ...09 2021 ECO Gold 85 134 Appendix with settings ...
Page 121: ...09 2021 ECO Gold 121 134 Click on the tab with the mouse and then on the Device manager ...
Page 129: ...09 2021 ECO Gold 129 134 ...
Page 130: ...130 134 ECO Gold 09 2021 ...
Page 131: ......