118 / 137
Proline Kryomats
Master: Alarm message
bl0C
Command: Pump blocked
(Pump motor monitoring: Overload, blockage).
1.
The viscosity of the heat teransfer liquid is too high
change heat transfer liquid or raise the setpoint
temperature.
2.
The pump is blocked.
Contact the LAUDA Service
Master: Alarm message
PulEU
Command: Low level (pump)
(Pump motor monitoring: No load).
1.
No liquid in the system. If this occurs, the level
monitoring has failed.
Check whether the float in the
expansion vessel is blocked by foreign bodies. Otherwise,
contact LAUDA Service
2.
With the option "open load" the device draws air out of
the open load.
Move the return to the load.
Master: Alarm message
Error 11
Command: Overpressure
(outflow pressure too high).
Pump level too high
Select a lower pump level
Compressor overtemperature
Three-phase current
9.4
Disposal information
The following applies for EU member states: The device must be disposed of according to
Directive 2012/19/EU (WEEE Waste of Electrical and Electronic Equipment).
9.4.1
Disposal of the refrigerant
Caution!
Uncontrolled escape of refrigerant / flying parts
Crushing, impact, cuts
No disposal with pressurized refrigerant circuit).
The decommissioning is only permitted by a specialist.
The type and filling quantity of the refrigerant can be read on the unit or on the rating plate. Repair and disposal only
through a qualified refrigeration engineer!
The following applies for EU member states: The disposal of the refrigerant must be carried out according to
EC Directive 2015/2067/EU in conjunction with 517/2014/EU.