29 / 137
Site the unit on a flat surface
The unit must not be put into operation if its temperature during
storage or transport has dropped below the dew point.
Wait for about one hour.
The unit may NEVER be overturned nor put upside down!
Do not cover the ventilation openings at the back of the control
head and on all sides of the lower section of the unit.
Leave at least 40 cm of free space on all sides.
For operation as bath thermostat, set the bypass valve to
internal (operation without external loads) (
Plug the mains-cable from the Master control element to the
refrigerator unit into the socket 52H on the back of the
Kryomat.
Plug the LiBus connector of the Command remote control into
the 70S socket and secure it.
Plug the LiBus cable from the refrigerator unit also into the
socket 70S on the back of the Master control element and
secure it.
Operation with external loads
(Circulating thermostat) continue at (
0).
Check whether the pump connectors at the side and back are fitted with sealing caps (
3) or that
hoses are fitted for external loads.
Using bath temperatures over 70 °C the supplied self-adhesive label
should be applied on
the bath at an easily visible point.
Do not carry out technical changes on the device!
The edge of the bath must not be drilled!
It is essential to keep within the permissible ambient temperature range (
11).
An increased ambient temperature or increased cooling water temperature reduces the cooling
capacity.
6.2
Filling and draining
Filling
Caution!
Contact with heat transfer liquid when filling / draining
Harmful when inhaled, damage to eyes and skin
Pay attention to the safety data sheet for the heat transfer
liquid.
Use CE gloves, protective clothing and eye protection during
physical contact with heat transfer liquid.
Avoid splashing the heat transfer liquid.
Make sure that the drain tap is closed before filling.