30 / 137
Proline Kryomats
Caution!
Use of unsuitable heat transfer liquid
Explosion, burns, scalds, fire
When selecting the heat transfer liquid, observe the
admissible temperature range.
Only use LAUDA heat transfer liquids.
Notice
Overfilling containers, spilling heat transfer liquid
Environmental hazard from leaking heat transfer liquid
Note the thermal volume expansion of the heat transfer
liquid.
Where necessary, consider the displacement volume of the
body being introduced.
Take the volume of external consumers into account.
Close the drain cock.
Carefully remove all residues of the previous heat
transfer liquid (blow dry and remove screw plugs!).
Maximum filling level is up to 10 mm below the top
edge of the bath. Overfilling leads to the display of
the warning
WXArn103
(
Changes in volume of the heat transfer liquid during
temperature changes should be taken into account.
Best operation is with a level 30-80 mm below the
top edge of the bath.
The cooling pipe of the evaporator should be
covered.
Low-level cut-off occurs at about 95 mm below the
top edge of the bath.
The units are designed for use with non-flammable and flammable liquids to DIN EN 61010-2-
010. Flammable heat transfer liquids may only be used below the flash point. (
When using heat transfer oils note that they expand on heating (approx. 10 %/100 K). With
enclosed external loads, the overall expansion takes place in the bath of the Proline Kryomat.
Ensure that with the connection of an external load, the liquid level does not drop impermissibly due
to filling the load
top up with liquid if necessary.
Set the upper and lower temperature limits (
7.6.2) in accordance with the limits of the heat
transfer liquid in use.