86 / 137
Proline Kryomats
The window shown adjacent appears.
With the menu
Status
the test run of the
Self Adaption can be started. When the Self
Adaption is finished, the test run will be
terminated automatically.
As soon as start
is pressed, in the sofkey
area the information
Adaption on
will be displayed followed by
the actual status of the test run.
With the menu
Setpoint
the set
temperature for the test run can be set.
The bath temperature will oscillate less than
about ±15 Kelvin around the set
temperature.
Change the display in the adjacent window
and accept with
.
With the menu
Identification
the optimal
control parameters for internal control or for
the internal control and the external control
can be detected automatically.
To detect the control parameters for the
external application, a temperature probe
must be connected to the thermostat.