Proline Kryomats
103 / 137
Command
Explanation
OUT_MODE_00_X
Keyboard Master: 0 = free / 1 = locked (corresponds to “KEY”).
OUT_MODE_01_X
Control: 0 = internal / 1 = external Pt100 / 2 = external analogue / 3 = external serial / 5 =
external Ethernet / 6 = external EtherCAT.
Note: With some temperature control devices this command can only be carried out if the
command OUT_PV_05_XXX.XX has been sent by the interface. Also notice that the
command OUT_PV_05_XXX.XX must be cyclically transmitted by the selected interface.
OUT_MODE_03_X
Keyboard Command remote control: 0 = free / 1 = locked.
OUT_MODE_04_X
Setpoint offset source: 0 = normal / 1 = external Pt / 2 = external analogue / 3 = external
serial.
Note: With some temperature control devices this command can only be carried out if the
command OUT_PV_05_XXX.XX has been sent by the interface. Also notice that the
command OUT_PV_05_XXX.XX must be cyclically transmitted by the selected interface.
START
Switches the unit on (after Standby). See safety information (
STOP
Switches the device into Standby (pump, heater, cooling unit OFF).
RMP_SELECT_X
Selection of the program (1 – 5) to which the further instructions apply. When the unit is
switched on, program 5 is selected automatically.
RMP_START
Start the programmer.
RMP_PAUSE
Hold (pause) the programmer.
RMP_CONT
Restart the programmer after pause.
RMP_STOP
Terminate the program.
RMP_RESET
Delete the program (all Segments).
RMP_OUT_00_XXX.XX_XXXXX_XXX.XX_X
Set a programmer segment (temperature, time, tolerance and pump level). A segment is
added and appropriate values are applied to it.
RMP_OUT_02_XXX
Number of times the program runs: 0 = unlimited / 1 – 250.
For ”_“ use also ” ” (blank character).
Response from thermostat ”OK“ or in case of error ” ERR_X“ (RS 485 interface e.g.
“A015_OK” or in case of error ”A015_ERR_X”.).
The command from the computer must be terminated with CR, CRLF, or LFCR.
The response of the thermostat is always terminated with CRLF.
After each command sent to the thermostat, it is necessary to wait for the reply before
sending another command. This ensures that the sequencing of inquiries and answers is
clear.
CR = Carriage Return (Hex: 0D)
LF = Line Feed (Hex: 0A)
Permitted data formats:
-XXX.XX
-XXX.X
-XXX.
-XXX
XXX.XX
XXX.X
XXX.
XXX
-XX.XX
-XX.X
-XX.
-XX
XX.XX
XX.X
XX.
XX
-X.XX
-X.X
-X.
-X
X.XX
X.X
X.
X
-.XX
-.X
.XX
.X