Proline Kryomats
111 / 137
9.1.5
Fault memory
For the analysis and localization of faults the Command version includes a fault memory in which up to 45 fault and
alarm messages are saved.
Command
Error store
Settings
Device status
Error
store
Display
.
The last message is at the top.
Each message line can be marked with the
cursor keys. The message appears in plain text
in the footer.
Under “Source”, the CAN node is displayed
which signaled the fault.
Code is the number, which in the Master is
shown in the display until the cause has been
rectified.
Type: Alarm, Warning or Fault (Error).
9.2
Cleaning
9.2.1
Cleaning the surface of the device
Warning!
Live parts in contact with cleaning agents
Electric shock hazard
Disconnect the device from the mains before cleaning.
Notice
Live parts in contact with cleaning agents
Property damage
Disconnect the device from the mains before cleaning.
Water and other liquids must not enter the control head.
Cleaning can be carried out with water to which a few drops of surfactant (washing-up liquid) have been added and
using a damp cloth.
Carry out appropriate decontamination if hazardous material is spilt on or in the equipment.
The cleaning or decontamination method is determined by the user's specialist knowledge. In case of
doubt, contact the manufacturer.
No. Source Code Type Date
Time
10 Safety 2 Alarm ------------ -----------
9 Safety 4 Warn. 28.08.03 15:32:02
8 Contro. 32 Error 17.07.0310.:52:02
7 Contro. 3 Warn. 06.06.03 11:15:11
6 Contro. 9 Alarm 05.06.03 08:45:01
5 Contro. 3 Alarm 01.06.03 17:58:22
4 Contro. 4 Warn. 28.05.03 20:01:22
3 Contro. 5 Warn. 27.05.03 07:58:00
Low level
Pump Menu
End
T
set
T
fix