Proline Kryomats
73 / 137
No.
Pump
Out 1
Out 2
Out 3
Start
-------
-------
-------
-------
1
4
-------
-------
-------
Pump
Menu
End
Insert
Delete
Then continue with
to the pump and signal
output setting.
The right-hand part of the entry table appears
as shown on the left.
Here, in the "Pump" field, the pump level and,
in the fields "Out 1" to "Out 3", the contact
outputs of the contact mode (accessory) can
be programmed. With the setting "-------"
the starting value is retained which was either
set before the program start or was defined by
a previous segment in the running program.
Further details are given on the following pages.
A new segment is produced by moving the cell
with the black background to a blank line with
the cursor keys and then pressing the soft key
Insert . The values of the cell located above
it are automatically copied.
If the field in the column
T end °C
has a black
background, the entry mode "End of segment
temperature" is obtained by pressing the
key. Depending on the setting, that is the
temperature, which the thermostat is to
achieve on the internal or external temperature
probe.
Enter the value, confirm with the
key and
continue to the "Time" entry field with
.
If the field in the column
Time
" has a black
background, the entry mode for the "Segment
time" time setting is obtained by pressing the
key.
If 0 is entered into the field "Time", “------”
appears. Then the final temperature is
approached as quickly as possible. With a time
entry the final temperature is obtained exactly
after the time expires (ramp).
Enter the segment time and confirm with the
key.
Continue to the "Tolerance" entry field with
.
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
End of segment temperature:
25,
00
Min: -150.00°C Max:450.00°C
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Input segment time:
003:00
Hours(max.999):Minutes