7.12.5
Pump-motor supervision: Overload or blockage
The SelfCheck Assistant monitors the Varioflex pump:
1.
Alarm sounds as dual-tone signal for pump-motor overload or blockage.
2.
Display of
bL0C
signals blockage.
3.
The red LED
above the fault triangle
flashes.
Heater switches off on both poles,
Pump and refrigerating unit are switched off electronically.
Find the cause of the fault. Perhaps the viscosity of the heat transfer liquid is
too high or the pump is blocked.
Press the Enter key.
Also press this key if the unit has been switched off in the fault state.
Command
Pump-motor alarm!
Pump-motor alarm is shown in the display and signifies that unlocking is
only possible on the Master control panel.
7.12.6
Pump-motor supervision: Dry running
The SelfCheck Assistant monitors the Varioflex pump:
1.
Alarm sounds as dual-tone signal when the pump runs without liquid. This
can only occur when the float level measurement has failed.
2.
The display of
PuLEU
signals that the SelfCheck Assistant has
detected a pump low level.
3.
The red LED
above the fault triangle
flashes.
Heater switches off on both poles,
Pump and refrigerating unit are switched off electronically.
The cause of the failure of the level measurement with the floatation sensor must be found and
rectified. Perhaps foreign bodies in the bath block it.
Press the Enter key.
Also press this key if the unit has been switched off in the fault state.
Command
Alarm! Low level (pump)
Alarm! Low level (pump) is shown in the display and signifies that unlocking
is only possible on the Master control panel.
Pump alarm
bL0C
°C
Pump alarm
PuLEU
°C