Proline Kryomats
31 / 137
Draining
Caution!
Contact with hot / cold heat transfer liquid
Scalds, frostbite
Bring heat transfer liquids to room temperature before
draining.
Make sure that the drain tap is closed after draining.
Caution!
Delay in boiling and thermal decomposition due to liquid residues
Burns, scalds, development of harmful vapors
Remove all old heat transfer liquid completely from the bath,
external consumers, accessories and hoses. Flush and clean
them with new heat transfer liquid.
Switch off the thermostat on the mains switch (rotary
switch) withdraw the mains plug.
The drain tap (1) is located behind the front panel.
Fit the hose onto the draining nozzle (2).
Open the drain tap and run off the heat transfer liquid.
Close the drain tap.
Follow the regulations for the disposal of used heat transfer liquids.
6.3
Connection of the cooling water
Note that the following conditions apply for the connection of the cooling water supply:
Cooling water pressure (feed - outlet)
maximum 10 bar overpressure
Differential pressure (feed - outlet)
minimum 3.0 bar
Cooling water temperature
10 to 15 °C recommended,
10 to 30 °C admissible (with power restrictions)
Cooling water quantity
see Technical Data (
Cooling water hose for connection to the device
minimum 13 mm