Lake Shore Model 475 Gaussmeter User’s Manual
Service
8-7
Firmware Replacement (Continued)
CAUTION:
The ICs are Electrostatic Discharge Sensitive (ESDS) devices. Wear shock-proof wrist straps
(resistor limited to <5 mA) to prevent injury to service personnel and to avoid inducing an
Electrostatic Discharge (ESD) into the device.
3. Use IC puller to remove existing IC from the socket.
4. Noting orientation of new IC, use an IC insertion tool to place new device into socket.
Eprom.eps
5.
Follow the top of enclosure
INSTALLATION
procedure in Paragraph 8.8.
8.10 CONNECTOR AND CABLE DEFINITIONS
The AUXILIARY I/O, PROBE INPUT, SERIAL I/O (DTE), and IEEE-488 INTERFACE connectors are defined in
Figures 8-4 thru 8-7.
Auxiliary_IO.bmp
Pin
Description
Pin
Description
1
2
3
4
5
6
7
8
9
10
11
12
13
Analog Out 1
Analog Out 2
Analog Out 3
No Connection
Trigger In
Trigger Out
No Connection
Relay 1 NO
Relay 1 COM
Relay 1 NC
Relay 2 NO
Relay 2 COM
Relay 2 NC
14
15
16
17
18
19
20
21
22
23
24
25
Ground
Ground
Ground
No Connection
No Connection
No Connection
No Connection
No Connection
No Connection
No Connection
No Connection
No Connection
Figure 8-4. AUXILIARY I/O Connector Details