Lake Shore Model 475 Gaussmeter User’s Manual
Operation
DC Operation
4-9
Example:
If the present maximum reading is +20 kG and the instrument measures –35 kG,
the –35 kG becomes the new maximum reading.
If the minimum reading is –3.0 kG and the instrument measures –1.5 kG,
the –1.5 kG becomes the new minimum reading.
0 kG
+20 kG
+30 kG
–35 kG
–3 kG
–1.5 kG
Original Max
Example of operation
Magnitude
New Max
Original Min
New Min
In Algebraic mode, the field values are compared mathematically with the sign included. The maximum reading is the
larger signed value since the last reset.
Example:
If the present maximum reading is –30 kG and the instrument me10 kG,
the +10 kG becomes the new maximum reading.
If the minimum reading is +3.0 kG and the instrument measures –15 kG,
the –15 kG becomes the new minimum reading.
0 kG
+3 kG
+10 kG
–30 kG
–15 kG
+30 kG
New Max
Example of operation
Algebraic
Original Max
New Min
Original Min
To configure the Max Hold measurement mode,
press and hold
the
Max Hold
key for approximately 4 seconds. The
max hold setup screen appears as a prompt for selecting the measurement mode.
Sel ec t Wi t h ° ®
Ma x Ma g n i t u d e
Use the
s
or
t
key to select Magnitude or Algebraic. Press
Enter
to accept the new selection and continue to the next
setting screen. Press
Escape
to cancel the new selection and return to the normal display.