Lake Shore Model 475 Gaussmeter User’s Manual
4-20
Peak Operation
Operation
4.8
PEAK MEASUREMENT MODE
To measure pulsed or periodic fields, press the
Peak
key on the front panel. In Peak measurement mode, the keypad and
functionality of the Model 475 is optimized to provide the best interaction for Peak measurements. The keypad features
are described in the following paragraphs.
NOTE
: Max Hold is not offered for Peak Measurement Mode.
4.8.1
Peak Operation Select Range
The Model 475 reads each Lake Shore probe type: High Stability, High Sensitivity, and Ultra-High Sensitivity. The
tables below list full scale ranges for each probe sensitivity, along with the display resolution. Measurement resolution
(noise floor) varies depending on probe and application.
High Stability Probe (HST)
Gauss
Tesla
Oersted
Amp/meter
Range and Resolution
Range and Resolution
Range and Resolution
Range and Resolution
±
350.00 kG
±
35.000 T
±
350.00 kOe
±
28.000 MA/m
±
35.000 kG
±
3.5000 T
±
35.000 kOe
±
2.8000 MA/m
±
3.5000 kG
±
350.00 mT
±
3.5000 kOe
±
280.00 kA/m
±
350.00 G
±
35.000 mT
±
350.00 Oe
±
28.000 kA/m
±
35.000 G
±
3.5000 mT
±
35.000 Oe
±
2.8000 kA/m
High Sensitivity Probe (HSE)
Gauss
Tesla
Oersted
Amp/meter
Range and Resolution
Range and Resolution
Range and Resolution
Range and Resolution
±
35.000 kG
±
3.5000 T
±
35.000 kOe
±
2.8000 MA/m
±
3.5000 kG
±
350.00 mT
±
3.5000 kOe
±
280.00 kA/m
±
350.00 G
±
35.000 mT
±
350.00 Oe
±
28.000 kA/m
±
35.000 G
±
3.5000 mT
±
35.000 Oe
±
2.8000 kA/m
±
3.5000 G
±
350.00 µT
±
3.5000 Oe
±
280.00 A/m
Ultra-High Sensitivity Probe (UHS)
Gauss
Tesla
Oersted
Amp/meter
Range and Resolution
Range and Resolution
Range and Resolution
Range and Resolution
±
35.000 G
±
3.5000 mT
±
35.000 Oe
±
2.8000 kA/m
±
3.5000 G
±
350.00 µT
±
3.5000 Oe
±
280.00 A/m
±
350.00 mG
±
35.000 µT
±
350.00 mOe
±
28.000 A/m
±
35.000 mG
±
3.5000 µT
±
35.000 mOe
±
2.8000 A/m
For manual ranging, press the
Select Range
key. The screen appears as a prompt for changing the range.
Sel ec t Wi t h ° ®
Ra n g e + 3 . 5 0 0 0 k G
Use the
s
or
t
key to select from the available ranges. Press Enter to accept the new selection and return to the normal
display. Press Escape to cancel the new selection and return to the normal display.
NOTE
: The Auto Range function is not offered for Peak measurements.