Lake Shore Model 475 Gaussmeter User’s Manual
6-38
Remote Operation
RDGMNMX?
Minimum and Maximum Reading Query
Input:
RDGMNMX? [term]
Returned:
<min>,<max>[term]
Format:
±
nnn.nnnE
±
nn,
±
nnn.nnnE
±
nn
Remarks:
Returns the most recent minimum and maximum field readings.
RDGOHM?
Resistance Reading Query
Input:
RDGOHM? [term]
Returned:
<hall resistance>[term]
Format:
±
nnn.nnnE
±
nn
Remarks:
Returns the Hall resistance of the sensor.
RDGPEAK?
Peak Reading Query
Input:
RDGPEAK? [term]
Returned:
<negative peak>,<positive peak> [term]
Format:
±
nnn.nnnE
±
nn,
±
nnn.nnnE
±
nn
Remarks:
Returns the negative and positive peak readings.
RDGREL?
Relative Reading Query
Input:
RDGREL? [term]
Returned:
<relative reading>[term]
Format:
±
nnn.nnnE
±
nn
Remarks:
Returns the relative field reading.
RDGTEMP?
Probe Temperature Reading Query
Input:
RDGTEMP? [term]
Returned:
<temperature>[term]
Format:
±
nnn.nnnE
±
nn
Remarks:
Returns the probe temperature reading in a format based on the present temperature units.
REL
Relative Mode Command
Input:
REL <off/on>, <setpoint source>[term]
Format:
n,n
<off/on>
Specifies Relative mode off or on: 0 = off, 1 = on.
<setpoint source> Specifies source of relative setpoint: 1 = User defined, 2 = Present Field.
Example:
REL 1,1 [term]
– Relative mode turned on, configured to use the User defined setpoint.
Refer to RELSP command.
REL?
Relative Mode Query
Input:
REL? [term]
Returned:
<off/on>, <setpoint source>[term]
Format:
n,n
(Refer to command for description)