Lake Shore Model 475 Gaussmeter User’s Manual
7-12
Probes and Accessories
7.8
ZERO GAUSS CHAMBER
NOTE:
Use care to ensure the Zero Gauss Chamber does not become magnetized. Using a magnetized
chamber to zero a probe can lead to erroneous field readings. It is a good practice to periodically
degauss the chamber. If no professional degausser is available, a bulk tape degausser (Verity VS250,
Data Devices PF211, or equivalent) may be used.
4060_Chamber.eps
Figure 7-17. Model 4060 Zero Gauss Chamber
4065_Chamber.eps
Figure 7-18. Model 4065 Large Zero Gauss Chamber