Lake Shore Model 475 Gaussmeter User’s Manual
4-18
RMS Operation
Operation
4.7.5
RMS Operation Max Hold
The Max Hold function captures the largest (Maximum) or smallest (Minimum) RMS field readings since the last
Reset
press. To turn the Max Hold feature on, press the
Max Hold
key. The following screen will appear for approximately 3
seconds.
Max Hol d On
To turn the Max Hold feature off, press the
Max Hold
key again.
4.7.5.1
Max/Min Display Setting
Typically, the maximum reading (Max) is displayed on the top line when the Max Hold feature is on. Sometimes it may
be beneficial to display the minimum reading (Min) on the top line instead of the maximum. The user may also display
both the maximum and minimum readings, with the maximum reading on the top line and the minimum reading on the
bottom line.
Displaying both will override the two line display configuration when Max Hold is active
(refer to
Paragraph 4.4.1).
To configure the Max Hold display setting, continue from the Max Hold mode screen or
press and hold
Max Hold
and
then press
Enter
until the following screen appears as a prompt for Max Hold display setup.
Sel ec t Wi t h ° ®
Ma x Di s p Ma x
Use the
s
or
t
key to select Max, Min, or Both. Press
Enter
to accept the new selection and return to the normal
display. Press
Escape
to cancel the new selection and return to the normal display.
4.7.6
RMS Operation Reset
If the Max Hold function is on and the
Reset
key is pressed, the maximum and minimum readings are cleared and reset
with the present field reading.
4.7.7
RMS Operation Relative
The relative function lets the user see small variations in larger fields. When the relative function is on, the relative
readings will appear on the top line of the display including the small delta sign (
s
) signifying the relative display. The
displayed reading is equal to the present field value minus the relative setpoint.
The relative setpoint can be configured to be the present field reading or it can be a user defined value. The relative
setpoint can be displayed on the bottom display. Refer to Paragraph 4.4.1.
To configure the relative setpoint,
press and hold
the
Relative
key for approximately 4 seconds. The relative setup
screen appears as a prompt for selecting the source of the relative setpoint.
Sel ec t Wi t h ° ®
SP Pr e s e n t Fi e l d
Use the
s
or
t
key to select Present Field or User Defined. Press
Enter
to accept the new selection and return to the
normal display. Press
Escape
to cancel the new selection and return to the normal display.