Lake Shore Model 475 Gaussmeter User’s Manual
6-32
Remote Operation
DFLT
Factory Defaults Command
Input:
DFLT 99 [term]
Remarks:
Sets all configuration values to factory defaults and resets the instrument. The "99" is included to
prevent accidentally setting the unit to defaults.
DISPLAY
Display Configuration Command
Input:
DISPLAY <item>[term]
Format:
n
<item>
Specifies the item to display on line 2: 1
–
6
1 = Blank
2 = Field
3 = Probe Temperature and Frequency (if available)
4 = Relative setpoint
5 = Field Control setpoint
6 = Alarm Sort message
Example:
DISPLAY 3[term]
– Probe Temperature will be displayed on Line 2 if the probe is capable of
measuring temperature. The Frequency will also be displayed if the Model 475 is in RMS mode.
DISPLAY?
Display Configuration Query
Input:
DISPLAY? [term]
Returned:
<item>[term]
Format:
n
(Refer to command for description)
DLOG
Datalog Stop/Start Command
Input:
DLOG <stop/start> [term]
Format:
n
<stop/start>
Starts and stops the datalog function: 0 = Stop, 1 = Start.
Remarks:
Use the DLOGSET command to set the datalog rate. The DLOGNUM? command can be used to
determine how many datalog points have been acquired.
DLOGNUM?
Datalog Number of Points Query
Input:
DLOGNUM? [term]
Returned:
<num rdgs>[term]
Format:
nnnn
Remarks:
Returns the number of field readings stored in the Datalog buffer.
DLOGRDG?
Datalog Reading Query
Input:
DLOGRDG? <data point>[term]
Format:
nnnn
<data point>
Specifies data point to retrieve: 1
–
1024.
Returned:
<value>[term]
Format:
±
nnn.nnnE
±
nn
Remarks:
Returns the field reading stored in the Datalog buffer at the point specified. Readings are in the field
units the instrument is currently set to.