![Lake Shore 475 User Manual Download Page 141](http://html.mh-extra.com/html/lake-shore/475/475_user-manual_670635141.webp)
Lake Shore Model 475 Gaussmeter User’s Manual
Service
8-13
17.
Configure the 475 range to the 3.5
Ω
range (RANGE 4).
18.
Read the resistor using the 475 (RDGOHM?). Store as ACTUALRESISTANCE.
19.
Calculate the Gain Calibration Factor (GCF) = EXPECTEDRESISTANCE /ACTUALRESISTANCE. Test this
factor to be 1.0, ±2%.
20.
Send the Gain Calibration Factor (GCF) to the Model 475 (CALG 2,4,<GCF>).
21.
Read the resistor using the 475 (RDGOHM?). Verify the reading to be EXPECTEDRESISTANCE ±0.02%.
22.
Repeat steps 14 to 21 for the 350 m
Ω
and 35 m
Ω
ranges (range 3 and 2) using the 300 m
Ω
and 30 m
Ω
resistors
respectively.
23.
Send the Gain Calibration Factor (GCF) calculated for range 2 to range 1. The hardware gains for those two ranges
are equivalent.
24.
Configure the 475 to use the 100 mA excitation current (HALLCS 1).
25.
Send the CALSAVE command to save the calibration constants into non-volatile memory.
8.11.2.3
Gaussmeter Calibration, 1mA Excitation Ranges
This section describes the method of calibrating the 1 mA current source that may be used for future Hall sensors. The
five ranges of the 475 will need to be calibrated for each current setting. The highest range is for each current is
calibrated using a 332
Ω
resistor that is measured using the 3458 in a four lead resistance measurement mode. The
remaining ranges are calibrated by using the previous calibrated range as shown in the following steps.
1.
Set the Gain Correction Factor to 1 for all five of the 1 mA hall current source ranges (CALG 3,1,1;CALG 3,2,1…).
2.
Connect the 332
Ω
resistor to the DVM using a proper 4-lead connection.
3.
Read the resistor using the DVM. Store as EXPECTEDRESISTANCE.
4.
Connect a short across the Gaussmeter input of the Model 475 using a proper 4-lead connection.
5.
Configure the 475 for 1 mA excitation current (HALLCS 3).
6.
Configure the 475 for DC 5 digits (RDGMODE 1,3,1,1,1) on the 350
Ω
range (RANGE 5).
7.
Zero the probe using the Zero Probe key or the ZPROBE command.
8.
Connect the 332
Ω
resistor across the Gaussmeter input of the Model 475 using a proper 4-lead connection.
9.
Configure the 475 for 1 mA excitation current (HALLCS 3).
10.
Read the resistor using the 475 (RDGOHM?). Store as ACTUALRESISTANCE.
11.
Calculate the Gain Calibration Factor (GCF) = EXPECTEDRESISTANCE/ACTUALRESISTANCE. Test this
factor to be 1.0, ±3%.
12.
Send the Gain Calibration Factor (GCF) to the Model 475 (CALG 3,5,<GCF>).
13.
Read the resistor using the 475 (RDGOHM?). Verify the reading to be EXPECTEDRESISTANCE ±0.02%.
14.
Connect the 33.2
Ω
resistor across the Gaussmeter input of the Model 475 using a proper 4-lead connection.
15.
Configure the 475 for 1 mA excitation current (HALLCS 3).
16.
Read the resistor using the 475 (RDGOHM?). Store as EXPECTEDRESISTANCE.
17.
Configure the 475 range to the 35
Ω
range (RANGE 4).
18.
Read the resistor using the 475 (RDGOHM?). Store as ACTUALRESISTANCE.
19.
Calculate the Gain Calibration Factor (GCF) = EXPECTEDRESISTANCE /ACTUALRESISTANCE. Test this
factor to be 1.0, ±2%.
20.
Send the Gain Calibration Factor (GCF) to the Model 475 (CALG 3,4,<GCF>).
21.
Read the resistor using the 475 (RDGOHM?). Verify the reading to be EXPECTEDRESISTANCE ±0.02%.
22.
Repeat steps 14 to 21 for the 3.5
Ω
and 350 m
Ω
ranges (range 3 and 2) using the 3
Ω
and 300 m
Ω
resistors
respectively.
23.
Send the Gain Calibration Factor (GCF) calculated for range 2 to range 1. The hardware gains for those two ranges
are equivalent.
24.
Configure the 475 to use the 100 mA excitation current (HALLCS 1).
25.
Send the CALSAVE command to save the calibration constants into non-volatile memory.