Lake Shore Model 475 Gaussmeter User’s Manual
6-40
Remote Operation
TUNIT
Probe Temperature Units Command
Input:
TUNIT <units>[term]
Format:
n
<units>
1 = Celsius, 2 = Kelvin
Example:
TUNIT 1[term] –
Configures the Model 475 to report probe temperature in °C.
TUNIT?
Probe Temperature Units Query
Input:
TUNIT? [term]
Returned:
<units>[term]
Format:
n
(Refer to command for description)
TYPE?
Probe Type Query
Input:
TYPE? [term]
Returned:
<type>[term]
Format:
nn
Remarks:
Returns the probe type:
40 = high sensitivity
41 = high stability
42 = ultra-high sensitivity
50 = user programmable cable/high sensitivity probe
51 = user programmable cable/high stability probe
52 = user programmable cable/ultra-high sensitivity probe
UNIT
Field Units Command
Input:
UNIT <units>[term]
Format:
n
<units>
1 = Gauss, 2 = Tesla, 3 = Oersted, 4 = Amp/meter.
Example:
UNIT 2[term] –
Configures the Model 475 to report readings in Tesla.
UNIT?
Field Units Query
Input:
UNIT? [term]
Returned:
<units>[term]
Format:
n
(Refer to command for description)
ZCLEAR
Clear Zero Probe Command
Input:
ZCLEAR [term]
Remarks:
Resets the value stored from the ZPROBE command.
ZPROBE
Zero Probe Command
Input:
ZPROBE [term]
Remarks:
Initiates the Zero Probe function. Place the probe in zero gauss chamber before issuing this command.