Maintenance and Service
NXT Go Owner’s Manual, LBL-00145, MAN2030
77
5.1.2.3
Roam NXT and PIM NXT Preparation
To prepare for cleaning or disinfection, remove the belt clips from the Roam NXT and the
PIM NXT. Refer to the
Removing and Attaching the Clip
on removing the clip.
When performing cleaning or disinfection, the Roam NXT and PIM NXT can be separated by
unfastening the screw holding the Roam NXT and the PIM NXT together. Disconnection of
the Roam NXT and PIM NXT is not required for cleaning or disinfection. For more
information on separating the Roam NXT and the PIM NXT refer to the
Identify the PIM NXT model in use. Ensure to close the transducer doors on the PIM NXT for
T-DOC.
IMPORTANT:
Do not attempt to clean the inside of the PIM NXT for T-DOC transducer
channels as this may damage the delicate measurement elements. This area should not
require cleaning as it is protected by the transducer doors while not in use or by the
catheter connector during Urodynamic study. If dirt or debris is observed in the transducer
channels as a part of daily inspection, Contact LABORIE for service and/or disposal
instructions.
If contamination is observed near the transducer channels during or post procedure, follow
the instructions provided in the
Cleaning the Roam NXT and PIM NXT
remove contamination from this area of the device.
Cleaning
Follow the instructions provided in this section to manually clean the NXT System and
peripherals.
IMPORTANT:
No method of automated cleaning or disinfection has been validated for
the NXT System or its modules. None of the NXT modules, Computer, or peripherals are
sterilizable or autoclavable. Refer to the manufacturer’s instructions for how to clean the
laptop computer.
5.1.3.1
Cleaning the Urocap NXT
This section provides instructions for cleaning the Urocap NXT and its accessories. Ensure
that the vent plug is inserted into the open vent hole, then follow the instructions provided
in the
to clean the Urocap NXT, being
certain not to apply excessive force to the top of the Urocap NXT.
IMPORTANT:
Excessive force applied to the top of the Urocap NXT, where the beaker
is placed, during cleaning or handling can damage the measurement apparatus.
If the area around the battery cover ventilation holes on the Urocap NXT becomes soiled,
this portion of the device should be cleaned as follows:
Remove the vent plug if connected.