Synergy Software
NXT Go Owner’s Manual, LBL-00145, MAN2030
68
4.5.4.4
Normal Values
Normal Values
is a data field available for activation within report customization. The
Normal Values
field is intended to provide an expected range of values for different events
which may occur during the study (Figure 57).
By default, the
Normal Values
field will be unchecked in the
Configuration
Settings
for the report.
To enable the
Normal Values
field,
a)
Select the required configuration copy, and then expand the
Report content
option by clicking the arrow beside the title:
.
b)
Click the checkbox beside the
Normal Values
field to enable this function.
c)
To save the changes to the report, click the
Apply Changes
button.
Figure 57: Normal Values
4.5.4.5
Sending Reports to EMR
Open the report from the
Patients
section or from the
Reports Workflow Step
and click
the
Send to EMR
button
on the report ribbon bar. This icon will only be visible if
the user has an EMR license enabled. When the
Send to EMR
button is clicked, Synergy
validates if the EMR service is running. If the service is running Synergy will send the
report to the EMR System. If the EMR service has stopped, Synergy will display a
notification message. All studies in Synergy can be sent to an EMR system if an EMR license
is enabled.
Synergy displays a progress indicator while waiting for acknowledgement from the EMR
system. Synergy will wait for 10 seconds, then display a message asking if the user wants
to continue waiting. This 10 second wait time is configurable and can be increased if
needed from Synergy application configuration file. After 60 seconds from the time the
report was sent to EMR, the process will timeout and show an
Export Failed
dialog if no
acknowledgement had been received from EMR system (Figure 58).