How to Perform a Study
NXT Go Owner’s Manual, LBL-00145, MAN2030
41
Select the
Stress Profile (On)
toggle button to start the stress profile. The UPP
Puller will begin pulling at the speed selected in the
Workflow Configuration
window.
Select the
Stress Profile (Off)
toggle button to mark the end of the profile. Click
the
Puller Return
button to return the UPP Puller NXT to the start position.
Manually reinsert the catheter and position just proximal to the bladder neck to
repeat the profile. Once the required number of profiles is complete, finish the
selected study.
3.6 Finishing the Studies
When all tests have been completed, remove any catheters and EMG patches from
the patient and dispose of them in accordance with hospital/clinic procedures.
Ensure to deflate Abdominal Fluid NXT Catheter Balloons before removal.
Empty the beaker. The beaker should then be thoroughly washed and reused.
Refer to the
Remove and dispose of the Infusion Pump Tubing and any other consumables
used during the study.
3.7 Printing the Study Result (Optional)
1.
Select the patient from the
Patients
section, on the main screen, and click the
Studies
tab in the
Patient Details
section.
2.
Click the
Report
button. The report window will then be displayed using default
configuration as shown in the figure below.
3.
Ensure that the printer is turned on and connected to the system. Click the
icon to display the
pop-up window and click the
OK
button to print the
report.
NOTE:
The user may also print the report at the end of the study once all data has
been recorded. A preview of the report is generated, the user can apply changes to the
report and then click the
icon.
3.8 Troubleshooting
For FAQs and troubleshooting tips, refer to the
. If problems cannot be resolved, contact the LABORIE Service team at 1-
Status DRAFT Effective