Maintenance and Service
NXT Go Owner’s Manual, LBL-00145, MAN2030
95
Symptoms
Possible Cause(s)
Corrective Action(s)
Software
Valsalva readings
are being marked
automatically as
coughs.
When performing
Valsalva procedure,
patient is triggering a
sharp rise in pressure
resembling a cough
spike?
Activate Bristol cough marker in the
workflow configuration.
Click on
Settings
>
Workflow
Settings
>
Phases
. Create a copy of
the workflow. In each study phase
scroll to the
Workflow Additional
Configuration Items
subtitle and de-
select
Auto Cough artifact
detection
.
Select
Auto Bristol artifact
detection
and save changes to custom
Workflow. Use the new custom
workflow to complete test with the
patient.
For more information on custom
workflows refer to the
Workflows section on page
If problems continue, contact the LABORIE Service team at 1-800-333-1039 or email
Software Error Messages
If the following error messages are displayed in Synergy software, contact the LABORIE
Service team at 1-800-333-1039 or email