How to Perform a Study
NXT Go Owner’s Manual, LBL-00145, MAN2030
37
Caution:
Infusion can be stopped at ALL TIMES by pressing the grey pump stop button
on the NXT Go Pump Hub, by opening the pump door, or by disconnecting the filling tube
from the catheter or the water bag.
3.3.2.4
Voiding Phase
The
Voiding
phase will begin (Figure 24). Click the
Permission To Void
button and ask
the patient to void through the commode chair into the Urocap NXT setup. See
Preparing the Urocap NXT for a reference image of the Urocap NXT Setup. If the
patient fails to void re-click the
Permission to Void
button to create a new
marker; this action will shift the previous marker to ‘Attempt to Void’.
Figure 24: Pressure Flow Voiding Phase
The voiding phase will stop 60 seconds after voiding ends.
A Post Voided Residual
window (Figure 21) will pop-up. Select the preferred
method of capturing Post Voided Residual (PVR), meaning the amount of fluid
remaining in the bladder at the completion of the voiding phase. Click
OK
to
move to the Report.
3.3.2.5
Urology Urodynamics Report
A preview of the Pressure Flow Report will be generated using the default (UDS)
configuration settings. The user can apply the default configuration for the report
or customize the report. For information on customizing the report, refer to the
Click the
Finish
button to close the report. The report will then appear under the
Studies
tab in the
Patient Details
section.
The operator may begin a secondary test with the same patient, under the
guidance of the overseeing physician, or continue to finish the study.