Table 14: UPP Puller NXT LED Signals
6.10
T-DOC Air-Charged Catheters
The T-DOC NXT Air-Charged Catheter series are part of the next-generation LABORIE
systems designed for use during Urodynamic studies (Figure 90).
Figure 90: T-DOC NXT Air-Charged Catheter
The T-DOC Air-Charged Catheters are single-use catheters intended for
Urodynamic pressure monitoring. T-DOC Air-Charged Catheters are provided
sterile and must not be re-used, reprocessed or resterilized.
The T-DOC Air-Charged Catheters feature a connector developed for use with the
PIM NXT.
The connector is inserted into the charging cavity on the PIM NXT, and a small
amount of air is pumped through the catheter lumen(s) into the air balloon located
at the distal end of the catheter. A tactile click indicates that the connector is
securely locked in the PIM NXT.
When external pressure is applied to the balloon, the pressure transducer registers
this pressure differential and sends the data to the Roam NXT. For information on
how to setup the T-DOC Air-Charged Catheters, refer to the
Consumables and Preparing the Patient
Battery Status
UPP Puller NXT Status
UPP Puller
NXT
Breathing Green: Sufficiently charged or
charging.
Breathing Green: Ready
Breathing Green - Fast: Sampling or
actuator is in motion.
Breathing Amber: Battery is low (less
than 60 minutes of operation).
Breathing Amber: Puller is not ready for
sampling or motion.
Breathing Blue: Firmware upgrade is in
progress.
Solid Red: Fault
Lack of illumination: Puller is sleeping or
shut off.
Lack of illumination: Empty/Sleep mode
(No declared battery capacity)
Status DRAFT Effective