5.1 Cleaning and Disinfection
General Cleaning and Disinfection
This section provides limitations and recommendations for the cleaning and disinfection
processes to be performed by the user on the standard modules of the NXT System. See
the
. Synergy is an application software. It
does not require reprocessing. The Bluetooth Dongle is not intended for reprocessing.
Surfaces of the standard equipment intended for reprocessing should be cleaned and/or
disinfected following the reprocessing procedures outlined in the following sections,
beginning on page 75:
General Cleaning and Disinfection
Cleaning and Disinfection Preparation
Completing Cleaning and Disinfection
5.1.1.1
Cleaning and Disinfection Reagents
The standard modules of the NXT System intended for reprocessing, excluding the laptop
PC, are compatible with the following cleaning and disinfection reagent:
Ammonia based wipes
CAUTION:
LABORIE has not evaluated the use of cleaning/disinfection reagents which
do not appear above. Use of unapproved reagents could damage the NXT system and
should be avoided. Do not use wipes containing bleach. For questions on the compatibility
of a reagent which is not listed above, contact LABORIE.
IMPORTANT:
Ensure the reagents are used for at least the contact time stated in the
reagent instructions. Do not use reagents for more than twice the stated contact time. Do
not allow reagents to pool on any part of the NXT System or in the charging cradles.
The manufacturer of the laptop PC has declared the device to be compatible with the
following reagent:
70% isopropyl alcohol (final concentration by volume) on a moistened microfiber
cloth.
Refer to the latest instructions provided on the manufacturer’s website for the laptop PC
prior to beginning cleaning:
https://www.dell.com/support/article/en-
ca/sln308919/guidance-for-keeping-your-dell-technologies-equipment-clean?lang=en
5.1.1.2
Cleaning and Disinfection Frequency
Cleaning
–
After each use, all devices should be visually inspected and cleaned to
remove any debris or soiling.
Status DRAFT Effective