Maintenance and Service
NXT Go Owner’s Manual, LBL-00145, MAN2030
80
Remove any visible soiling by hand immediately after the completion of a study
using a dry, soft, lint free cloth.
Gently wipe down the surface of the Computer, NXT module, or
peripherals using a fresh cleaning wipe to remove visible soil.
Use a fresh wipe to thoroughly wet the external surfaces of the device. Wipe
around cables, edges, cracks, crevices, corners, and hard-to-reach areas. Ensure
that all surface areas and connectors have been in contact with the wipe.
Repeat as necessary until surfaces are visibly clean. If visible soil persists, use a
soft bristle brush to gently brush away or loosen the remaining soil, then wipe the
device again using a fresh wipe.
Wipe the modules, and peripherals with a dry, lint free cloth. Ensure the device is
dry to the touch before using the device in a procedure.
Visually inspect the units in standard office lighting to ensure they are clean. If the
inspection determines the devices to not be visually clean, the user should repeat
the cleaning cycle.
Perform visual inspection of all units for corrosion, discoloration, pitting, or
cracking. Contact LABORIE for service and/or disposal instructions if the unit is
observed with these defects.
NOTE
: There are no rinsing steps required for the cleaning or disinfection processes.
Disinfection
5.1.4.1
Performing Manual Disinfection
To disinfect the device, follow the cleaning instructions allowing the surface to
remain wet for the required contact time as listed on the product label of
the cleaning solution. Ensure to refer to module specific disinfection instructions.
Ensure the device is dry to the touch before using it to perform another study.
Once disinfection is complete, confirm that each module is working correctly by
verifying calibration. Refer to the
calibration instructions per module.
IMPORTANT
: Refer to Urocap NXT specific disinfection instructions in the
. Refer to Roam NXT and PIM NXT specific disinfection
Disinfecting the Roam NXT and PIM NXT
. Refer to the
Disinfecting the UPP Puller NXT
5.1.4.2
Disinfecting the Urocap NXT
To disinfect the Urocap NXT, follow the instructions provided in the
allowing the surface to remain wet for the required contact
time listed on the product label of the cleaning solution.
Disinfect the area around the battery cover ventilation holes on the Urocap NXT following
the instructions below:
Remove the vent plug if connected.
Gently wipe the affected area with a cleaning wipe to remove soiling and debris.
Take care to prevent any excess cleaning reagent from entering the vent hole.
Allow the surfaces to remain wet for the required contact time listed on the
product label of the cleaning solution.