OPERATION
6
55
WATERFLUX 3070
www.krohne.com
06/2016 - 4004999701 - MA WATERFLUX 3070 - R01 en
44
Status output 1:
P/T warning
0 (default)
Off
1
Pressure limit max. [menu item 62]
2
Pressure limit min. [menu item 63]
3
Temperature limit max. [menu item
67]
4
Temperature limit min. [menu item
68]
50
Status output 2:
Self check
0 Off (default)
1= On
51
Status output 2:
Battery pre warning
0 = Off (default)
Active in case of low battery
(remaining capacity < 10%).
1 = On
52
Status output 2:
Battery final warning
0 = Off (default)
Active in case of very low battery
(remaining capacity < 1%).
1 = On
53
Status output 2:
Empty pipe
0 = Off (default)
Active in case of empty pipe
detection.
1 = On
54
Status output 2
P/T warning
0=[default]
Off
1
Pressure limit max. [menu item 62]
2
Pressure limit min. [menu item 63]
3
Temperature limit max. [menu item
67]
4
Temperature limit min. [menu item
68]
No.
Function
Options
Description
WARNING!
When setting the values of the temperature and / or pressure, the minimum value should not
exceed the maximum value