6.2.6 Reset full access to the menu
To enable access to the menu again, follow the next steps:
•
Remove the cover.
•
Use a small screwdriver to press the reset button.
First press the reset button
1
and then simultaneously press both the optical keys for 6
seconds
2
.
•
The display will automatically switch to menu number 1.
•
To change the setting of menu number 1 from 1 to 0, please refer to
View or change menu
settings
on page 46.
•
For closing the case of the converter in the IP68 housing, please refer to
Closing of the
converter housing
on page 25.
WARNING!
Do not break the metrological seals if the meter has been verified to MID Annex III (MI-001) or
OIML R49 and has received metrological sealing. This can lead to the obligation to re-verify the
meter. Please check your local regulations.
Figure 6-6: Change the settings of the access control.
1
2
WARNING!
Make sure that the battery cable is not jammed by the cover.