3.8.3 Mounting of the KROHNE FlexPower
The mounting of the KROHNE FlexPower unit is possible in the following options:
•
Surface mounting with 2 screws on a wall or other sufficient hard surface
•
On-pipe mounting with 2 Tie-Wraps
For mounting on a horizontal or vertical surface, always use the proper tools and mounting
materials ( eg. drill , plug and screws). The distance between the two mounting holes is 184 mm
/ 7.2". Always mount the KROHNE FlexPower unit on the intended location before connecting to
the sensor and/or switching on the main power. When attaching the wall holder unit to the wall,
do not exceed a tightening torque of 1 Nm when fastening the screws. This can damage the
mounting eyelets.
Mounting on the pipe line construction can easily be done by using 2 Tie-Wraps. The bottom of
the housing of the KROHNE FlexPower unit is designed for mounting on a pipe line. Choose the
correct size and properties of the Tie-Wraps ( specifications according ambient temperature and
other conditions, size, width max.14 mm /
½
" ). Consider releasable types of Tie-Wraps when
mobility and/or replacing of mounting is expected.
The KROHNE FlexPower unit is IP68 rated.
CAUTION!
Prevent water ingress to the AC or DC cable when not connected.
Figure 3-20: dimensions KROHNE FlexPower
1
distance = 184 mm / 7.2".