E-18
Preparation
Inserting the Batteries
To supply power to the instrument, the AC adapter (AC-A17) or four AA-size batteries (Alkaline or Ni-
MH battery is recommended for better service life) must be used. Use either the AC adapter or batteries,
according to which suits your application.
N
ote
• If you are not going to use the instrument for more than two weeks, make sure that the batteries are removed.
If the batteries are left in the instrument for long periods of time, battery electrolyte may leak and damage the
instrument.
• Do not use batteries of different types or mix new batteries with old ones. Doing so may result in battery explo-
sion or reduction of battery life.
• Do not touch or short-circuit the terminals inside the battery chamber. Doing so may result in breakdown of
the instrument.
[Operating Procedure]
1
Make sure that power is OFF (i.e. the POWER
switch is set to “
”).
2
Slide the battery cover on the rear of the instru-
ment to open it.
3
Place four AA-size batteries in the battery
chamber. Make sure that the batteries are
placed in the correct direction.
4
Align the marker on the battery cover with that
on the instrument, and slide the battery cover
to close it.
Summary of Contents for CM-2600d
Page 1: ...Spectrophotometer CM 2600d 2500d E Instruction Manual Es Manual de instrucciones...
Page 11: ...E 9 Chapter 1 Before Using the Instrument...
Page 27: ...E 25 Chapter 2 Preparation for Measurement...
Page 62: ...E 60 Setting Color Difference Tolerances...
Page 63: ...E 61 Chapter 3 Measurement...
Page 79: ...E 77 Chapter 4 Other Functions...
Page 108: ...E 106 TASK Mode...
Page 109: ...E 107 Chapter 5 Troubleshooting...
Page 114: ...E 112 Troubleshooting...
Page 115: ...E 113 Chapter 6 Appendix...
Page 123: ...E 121 Dimensions...
Page 126: ......
Page 127: ...E Es 2002 2005 KONICA MINOLTA SENSING INC AFMAGA 1 Printed in Japan 9222 1856 40...