E-46
3
Turn
B
to select “ZERO”.
4
Direct the specimen measuring port into the air.
1m or more
• Never place the specimen measuring port toward a light source.
•
Keep the specimen measuring port more than 1m away from any reflective items (hands, desks,
walls etc.)
.
5
Make sure that
is displayed, then press
C
(MEAS. button).
Zero calibration will be performed.
When zero calibration is complete,
the <CALIBRAITION> screen will
reappear.
M
emo
• Use of the optional Zero Calibration Box ensures proper zero calibration.
If Zero Calibration Box is used, remove
“Measuring Base”.
• If
does not appear due to the auto power save function, this may be due the fact that it can take time
to start zero calibration after
C
is pressed.
N
ote
• When zero calibration is complete, perform white calibration. Measurement cannot be performed if
white calibration has not been completed.
• Even though white calibration has been performed after the power is turned ON, if zero calibration has
also been performed after that, white calibration must be performed again.
Summary of Contents for CM-2600d
Page 1: ...Spectrophotometer CM 2600d 2500d E Instruction Manual Es Manual de instrucciones...
Page 11: ...E 9 Chapter 1 Before Using the Instrument...
Page 27: ...E 25 Chapter 2 Preparation for Measurement...
Page 62: ...E 60 Setting Color Difference Tolerances...
Page 63: ...E 61 Chapter 3 Measurement...
Page 79: ...E 77 Chapter 4 Other Functions...
Page 108: ...E 106 TASK Mode...
Page 109: ...E 107 Chapter 5 Troubleshooting...
Page 114: ...E 112 Troubleshooting...
Page 115: ...E 113 Chapter 6 Appendix...
Page 123: ...E 121 Dimensions...
Page 126: ......
Page 127: ...E Es 2002 2005 KONICA MINOLTA SENSING INC AFMAGA 1 Printed in Japan 9222 1856 40...