E-86
Assessments
5
If they are satisfactory, turn
B
to select “BREAK”, then press
A
.
The measurement screen will ap-
pear.
6
Place the specimen measuring port to the specimen, and then press
C
(MEAS.
button).
The color difference from the target color, pass/fail judgment result based on the specified box tolerances
and direction of tint and brightness deviation will be displayed.
• The color space to be used as the assessment axis differs depending on the selected color space.
When “L*C*h, E”, “L*C*h, CMC1:1”, “L*C*h, CMC2:1” or “L*C*h, CIE94” is selected, assess-
ment will be made based on the direction of tint and lightness deviation and color difference from the
target color in the L*C*h axis.
When a color space other than those given above is selected, assessment will be made based on the
direction of tint and lightness deviation and color difference from the target color in the L*a*b* axis.
M
emo
• Assessment results differ as shown in the table below depending on the type of specified color difference toler-
ances and pass/fail judgment result.
Type
Specular
component
Displayed value
Pass/fail
judgment
Display content
Box
tolerance
I+E
Color difference
PASS
Measured value, with “PASS” to the right
FAIL
Highlighted measured value, with highlighted
message to the right
E*
PASS
Measured value, with “PASS” to the right
FAIL
Highlighted measured value, with highlighted
“FAIL” to the right
SCI or SCE
Color difference
PASS
Measured value, with “PASS” to the right.
FAIL
Highlighted measured value, with highlighted
“FAIL” and message to the right
E*
PASS
Measured value, with “PASS” to the right
FAIL
Highlighted measured value, with highlighted
“FAIL” to the right
Elliptical
tolerance
I+E
SCI or SCE
Color difference No judgment
Measured value, with message to the right
E*
PASS
Measured value, with “PASS” to the right
FAIL
Measured value, with highlighted “FAIL” to the
right
Summary of Contents for CM-2600d
Page 1: ...Spectrophotometer CM 2600d 2500d E Instruction Manual Es Manual de instrucciones...
Page 11: ...E 9 Chapter 1 Before Using the Instrument...
Page 27: ...E 25 Chapter 2 Preparation for Measurement...
Page 62: ...E 60 Setting Color Difference Tolerances...
Page 63: ...E 61 Chapter 3 Measurement...
Page 79: ...E 77 Chapter 4 Other Functions...
Page 108: ...E 106 TASK Mode...
Page 109: ...E 107 Chapter 5 Troubleshooting...
Page 114: ...E 112 Troubleshooting...
Page 115: ...E 113 Chapter 6 Appendix...
Page 123: ...E 121 Dimensions...
Page 126: ......
Page 127: ...E Es 2002 2005 KONICA MINOLTA SENSING INC AFMAGA 1 Printed in Japan 9222 1856 40...