E-118
Specifications
Specifications
Model
CM-2600d
CM-2500d
Illuminating/viewing system
d/8 (diffused illumination, 8-degree viewing angle), equipped with
simultaneous measurement (not switched mechanically) of SCI
(specular component included)/SCE (specular component exclud-
ed). (Conforms to CIE No.15, ISO 7724/1, ASTM E1164, DIN 5033
Tei17, JIS Z8722 Condition c, standard.)
Size of integrating sphere
ø52 mm
Detector
Silicon photodiode array (dual 40 elements)
Spectral separation device
Diffraction grating
Wavelength range
360 nm to 740 nm
Wavelength pitch
10 nm
Half bandwidth
Approx. 10 nm
Reflectance range
0 to 175%, Display resolution: 0.01%
Light sources
3 pulsed xenon lamps
2 pulsed xenon lamps
Measurement time
Approx. 1.5 seconds (approx. 2
seconds for fluorescent measure-
ment)
Approx. 1.5 seconds
Minimum measurement inter-
val
3 seconds for SCI/SCE (4 sec-
onds for fluorescent measure-
ment)
3 seconds for SCI/SCE
Battery performance
Approx. 1000 times at 10-second intervals
(when alkaline batteries used; at 23°C)
Measurement/illumination
area
MAV: ø8 mm/ø11 mm
SAV: ø3 mm/ø6 mm
(Selectable between MAV and
SAV)
MAV:ø8 mm/ø11 mm
Repeatability
Spectral Reflectance: Standard deviation within 0.1% (360 to 380
nm: Standard deviation within 0.2%)
Chromaticity Value: Standard deviation E*ab within 0.04 (When
a White Calibration Plate is measured 30 times at 10-second inter-
vals after white calibration)
Inter instrument agreement
E*ab within 0.2 (MAV/SCI) (Averaged for 12 BCRA Series II
color tiles compared to values measured with master body; at 23°C)
UV adjustment
Instantaneous numerical adjust-
ment (no mechanical adjustment
required)
* With UV400 nm cut filter
No UV adjustment function for
CM-2500d
Language modes
English/Japanese/German/French/Spanish/Italian
Measurement mode
Single/averaging (auto mode: 3, 5, 8 times/manual mode)
Interface
Conforms to RS-232C
Observer
2/10 degrees
Observation illuminant
A, C, D
50
, D
65
, F2, F6, F7, F8, F10, F11, F12 (simultaneous evalua-
tion is possible using two light sources)
Summary of Contents for CM-2600d
Page 1: ...Spectrophotometer CM 2600d 2500d E Instruction Manual Es Manual de instrucciones...
Page 11: ...E 9 Chapter 1 Before Using the Instrument...
Page 27: ...E 25 Chapter 2 Preparation for Measurement...
Page 62: ...E 60 Setting Color Difference Tolerances...
Page 63: ...E 61 Chapter 3 Measurement...
Page 79: ...E 77 Chapter 4 Other Functions...
Page 108: ...E 106 TASK Mode...
Page 109: ...E 107 Chapter 5 Troubleshooting...
Page 114: ...E 112 Troubleshooting...
Page 115: ...E 113 Chapter 6 Appendix...
Page 123: ...E 121 Dimensions...
Page 126: ......
Page 127: ...E Es 2002 2005 KONICA MINOLTA SENSING INC AFMAGA 1 Printed in Japan 9222 1856 40...