E-115
Principles of Measurement
Illumination Area and Measurement Area
The CM-2600d allows selection of either SAV (ø3 mm) or MAV (ø8 mm) measurement area according
to the specimen and applications. A Target Mask (illumination area) suitable for the selected measure-
ment area must be installed.
N
ote
With the CM-2500d, only MAV (ø8 mm) is available.
Target Mask
Since this instrument does not have automatic Target Mask matched to the selected measurement area
must be attached correctly.
Since the condition of the inner surface of the Target Masks affects measured values, do not touch the
inner surface by hand, scratch it or allow it to get dirty.
N
ote
With the CM-2500d, it is not possible to exchange a Target Mask since only MAV (ø8 mm) is available.
Measurement Area
The measurement area can be switched between SAV and MAV by sliding the measurement area selec-
tor provided on the side of the instrument.
N
ote
With the CM-2500d, no measurement area selector is provided.
Summary of Contents for CM-2600d
Page 1: ...Spectrophotometer CM 2600d 2500d E Instruction Manual Es Manual de instrucciones...
Page 11: ...E 9 Chapter 1 Before Using the Instrument...
Page 27: ...E 25 Chapter 2 Preparation for Measurement...
Page 62: ...E 60 Setting Color Difference Tolerances...
Page 63: ...E 61 Chapter 3 Measurement...
Page 79: ...E 77 Chapter 4 Other Functions...
Page 108: ...E 106 TASK Mode...
Page 109: ...E 107 Chapter 5 Troubleshooting...
Page 114: ...E 112 Troubleshooting...
Page 115: ...E 113 Chapter 6 Appendix...
Page 123: ...E 121 Dimensions...
Page 126: ......
Page 127: ...E Es 2002 2005 KONICA MINOLTA SENSING INC AFMAGA 1 Printed in Japan 9222 1856 40...