15
(4) Screw holes
There are two types of screws for securing a probe:
Tripod mount screw: Used for mounting the probe on a tripod. The depth is 6 mm.
ISO screw: Used for securing the probe to jigs. Use an ISO 5 mm screw and the depth is 6 mm.
•
The distance from each screw to the end of the optical column is the same as that for a CA-310 series probe. (Excluding
mini probes)
•
The position of the optical axis (in the height direction) is the same as that for a CA-310 series probe.
(5) Status indicator
Located on the rear of all probes to indicate the status of the probe.
No power supply: Off
During measurement or zero calibration: Off
Memory error: Blinking (On for 0.2 sec, then Off for 0.2 sec)
Zero calibration not performed: Blinking (On for 1 sec, then Off for 0.2 sec)
Measurement possible: On (even when measurement is not possible due to the setting of an invalid synchronization
frequency or integration time in INT, EXT, or MANUAL sync mode, without displaying that state)
Pr
obe Guide