64
Turning the Power ON/OFF
2. Instrument Status at Power-ON
By default, the instrument is in the following status when the POWER switch is set to ON.
1
Color Mode
Lvxy
2
Calibration channel No.
P. 82
CH 00
3
Target Value
P. 96
x = 0.3127, y = 0.3290, Lv = 100.0 (cd/m
2
)
4
PROBE
P. 79
P1
5
SYNC (measurement
synchronization mode)
P. 70
UNIV.
6
ID
P. 92
CH 00
7
Measurement speed
P. 68
AUTO
8
RS-232C baud rate
38400 bps
9
Correction coefficient for user
calibration
P. 83
Konica Minolta’s standard data
To Change the Instrument Status at Power-ON
(only for (1), (2), (4), and (5))
Change the necessary parameters and press the [ENTER] key for more than five seconds. A beep will sound, followed by a
long beep when the settings are saved. The instrument will start with the new settings when the power is turned ON next
time. (The selected mode and number will be stored in the memory, and they will remain effective even if the POWER switch
is set to OFF.) * For details, refer to the pages given in the above table.
Changing Method for (1) and (2)
(1) Color Mode ................... Press the
MODE
MODE key.
(2) Calibration channel ....... Press the [CAL CH] key followed by the [UP] or [DOWN] key.
Changing Method for (3)
(3) Target value ................... The current target value will be changed if a mode other than flicker is selected and a target
value is entered.
Changing Method for (4) to (9)
For (4) to (9), press the [MENU] key to display the menu screen, then switch to a relevant screen for setting.