112
Auto 0-Cal Settings
This instrument can be configured to perform zero calibration automatically.
When set to [ON], the shutter of the probe automatically closes and Auto 0-cal is performed if the probe temperature has
changed 6°C.
* Factory setting: OFF
Operating Procedure
1
While the measurement screen is
displayed, press the [MENU] key.
The setting screen is displayed on the LCD screen.
2
Press the [UP] or [DOWN] key to select
[Option], and then press the [RIGHT] or
[ENTER] key.
MENU
Trigger mode 0-cal AUTO
Meas Cond
User Calib
Target
Option
System
Absolute
3 (Normal)
ON
ON
AUTO
cd/
㎡
Color Space
Absolute/difference
Brightness
Buzzer
Auto 0-cal
Save Destination
Luminance Units
3
Press the [UP] or [DOWN] key to select
[Auto 0-cal], and then press the [ENTER]
key.
MENU
Trigger mode 0-cal AUTO
Meas Cond
User Calib
Target
Option
System
Absolute
3 (Normal)
ON
ON
AUTO
cd/
㎡
Color Space
Absolute/difference
Brightness
Buzzer
Auto 0-cal
Save Destination
Luminance Units
4
The screen for selecting whether to
perform Auto 0-cal is displayed. Press
the [UP] or [DOWN] key to select [ON] or
[OFF]
.
MENU
Trigger mode 0-cal AUTO
Meas Cond
User Calib
Target
Option
System
◉
ON
○
OFF
Color Space
Absolute/difference
Brightness
Buzzer
Auto 0-cal
Save Destination
Luminance Units
5
Press the [ESC] key.
The measurement screen is displayed on the LCD
screen.
The Auto 0-cal setting will be kept even if the POWER
switch is set to OFF (
).