36
CA-VP427
CA-VP410
Measurement area
Ø27 mm
Ø10 mm
Acceptance angle
±2.5°
±8.5°
Accuracy guaranteed measurement distance
30±10 mm
30±5 mm
Display range Luminance
0.0001 to 3000 cd/m
2
0.0001 to 3000 cd/m
2
Chromaticity
Displayed in 4 digits
Displayed in 4 digits
Luminance
Accuracy guaranteed luminance range
0.001 to 3000 cd/m
2
0.001 to 3000 cd/m
2
Accuracy (for white)
*1, *3
0.001 - cd/m
2
±9%
±9%
0.01 - cd/m
2
±2%
±2.5%
0.1 - cd/m
2
±1.5%
±2%
1 - cd/m
2
±1.5%
±2%
10 - cd/m
2
±1.5%
±1.5%
100 - cd/m
2
±1.5%
±1.5%
Repeatability
(2
σ
)
*1
AUTO
0.001 - cd/m
2
10%
7%
0.01 - cd/m
2
1%
1%
0.1 - cd/m
2
0.25%
0.25%
1 - cd/m
2
0.10%
0.10%
10 - cd/m
2
0.10%
0.10%
100 - cd/m
2
0.10%
0.10%
Chromaticity Accuracy guaranteed luminance range
0.01 to 3000 cd/m
2
0.01 to 3000 cd/m
2
Accuracy (for white)
*1, *3
0.01 - cd/m
2
±0.003
±0.003
0.1 - cd/m
2
±0.002
±0.002
1 - cd/m
2
±0.002
±0.002
10 - cd/m
2
±0.002
±0.002
100 - cd/m
2
±0.002
±0.002
In 100 cd/m
2
(for monochrome)
*2
100 cd/m
2
±0.003
±0.003
Repeatability
(2
σ
)
*1
AUTO
0.01 - cd/m
2
0.0030
0.0020
0.1 - cd/m
2
0.0008
0.0008
1 - cd/m
2
0.0003
0.0003
10 - cd/m
2
0.0002
0.0002
100 - cd/m
2
0.0002
0.0002
Flick
er (CA
-310 Mode)
*6
Flicker
(Contrast)
Measurement luminance range
---
---
Range to be measured (flicker frequency)
---
---
Accuracy
30 Hz, AC/DC 10% sine wave ---
---
60 Hz, AC/DC 10% sine wave ---
---
Repeatability (2
σ
)
20 to 65 Hz,
AC/DC 10% sine wave
---
---
Flicker
(JEITA)
Measurement luminance range
---
---
Range to be measured (flicker frequency)
---
---
Accuracy
30 Hz, AC/DC 4% sine wave ---
---
30 Hz, AC/DC 1.2% sine wave ---
---
Repeatability (2
σ
)
30 Hz, AC/DC 4% sine wave ---
---
30 Hz, AC/DC 1.2% sine wave ---
---
Specifications of Each Probe (4)
Summary of Contents for CA-410 Series
Page 12: ......
Page 40: ......
Page 139: ......
Page 140: ...En 9222 AA1J 12 2017 KONICA MINOLTA INC BIFBDA...