The screen for selecting a color is
displayed. Press the [UP] or [DOWN] key
to select the color to measure and then
press the [ENTER] key.
Complete
User Calib > RGB+W matrix calibration
Select Color
Measured Value
Calibration Value
01 CH01
MENU
8
When the pop-up screen for RGB+W
matrix calibration is displayed, press the
[ENTER] key to perform measurement.
When the measured value is displayed,
press the [ENTER] key to hold the value.
Complete
User Calib > RGB+W matrix calibration
Select Color
Measured Value
校正値
01 CH01
MENU
RGB+W matrix calibration
White
Start measurement
Meas
Measured Value
Target Value
Lv x y
9
Press the [UP] or [DOWN] key and the
[RIGHT]/[LEFT] keys to select the field of
a calibration value for user calibration,
and then press the [ENTER] key.
Complete
User Calib > RGB+W matrix calibration
Select Color
Measured Value
校正値
01 CH01
MENU
160.2 0.3132 0.3157
RGB+W matrix calibration
White
Stop measurement
Stop
Measured Value
Target Value
Lv x y
10
Press the [UP] or [DOWN] key and the
[RIGHT]/[LEFT] keys to set a calibration
value, and then press the [ENTER] key.
Press the [ESC] key to cancel configuration of the
setting and restore the original value.
11
Repeat steps 9 and 10 to set the
calibration value, move to [Complete],
and then press the [ENTER] key to
confirm your entry.
Complete
User Calib > RGB+W matrix calibration
Select Color
Measured Value
校正値
01 CH01
MENU
160.2 0.3132 0.3257
160.2 0.3132 0.3157
RGB+W matrix calibration
White
Edit Value
To confirm by pressing the ENTER button in the “Comp”
Complete
Measured Value
Target Value
Lv x y
Complete
User Calib > RGB+W matrix calibration
Select Color
Measured Value
Calibration Value
01 CH01
MENU
160.2 0.3132 0.3257 160.0 0.3130 0.3260
12
Repeat steps 7 to 11 to set calibration
values for W, R, G, and B, move to
[Complete], and then press the [ENTER]
key to confirm your entry.
(If the entered value is out of the available setting
range, the screen returns to the value entry screen.
Enter another value.)
13
Press the [ESC] key.
The
measurement screen
is displayed on the LCD
screen.
The specified values will be kept even if the POWER
switch is set to OFF (
).
CA
-DP40
Calibr
ation/
Set
tings