22
Probes
•
For RS-232C connection with a PC or PLC, connect the mini DIN plug of an optional conversion cable IF-A29 to the
RS-232C connector of the probe, and the D-sub plug to the RS-232C connector of the PC or PLC via a cross cable. The
baud rate of the probe is set to 38,400 bps.
In this case, it is also necessary to use the cable's power supply line to supply power to the probe. As a power source,
5 V/300 mA is required.
To control a probe from a PC via RS-232C, use the supplied SDK (CA-SDK2). For the specifications of the SDK and how to
use it, refer to the instruction manual for CA-SDK2.
RS-232C connection (using IF-A29)
+ −
Mini DIN plug
RS-232C connector
IF-A29
Cross
connection
BNC connector
External synchronization signal
5 V DC/300 mA
Red
Black
PC
RS-232C connector
RS-232C connector pin No. and functions
Mini-DIN 8-pin (female)
Pin No.
Pin function
Details
1
ExSync or Trigger IN
0 to 5 V, CMOS level, Max. 240 Hz
2
Trigger OUT
3
External power supply (5 V)
Min. 4.9 V to Max. 5.5 V
4
TxD
5
GND
6
RTS
7
RxD
8
CTS
Trigger timing diagram
Trigger IN
Trigger OUT
Transition to trigger mode
Measurement start Measurement end
Measurement start Measurement end
t > 5 μsec
t > 5 μsec
•
To input trigger signals or external synchronization signals to the probe, input the signals to the BNC plug of an optional
conversion cable IF-A29 or an optional BNC conversion cable IF-A35.
Input to the BNC plug
C-MOS logic level
BNC connector
Input trigger and external
synchronization signals
(Luminance/chromaticity measurement: 0.5 to 240 Hz)
(Flicker measurement: 0.5 to 130 Hz)
(Operated
with 5 V)
Summary of Contents for CA-410 Series
Page 12: ......
Page 40: ......
Page 139: ......