5-4
7F3319
October 4, 2004
Optimizing Image Quality
Table 1. Selection of Parameters by Modality Type
Modality Type
TFT Set
Table No.
(Contrast
Setting)*
Image
Density
(Dmax)*
Smooth-
ing Type
Note
Kodak
and other
CR
(Computed Radiography)
and
DR
(Digital
Radiography) Systems.
Also Workstations and
imaging devices that output
data to a linear-in- density
contrast curve.
WRKSTN2A.w87
1
2
3
4
3.1
3.0
2.9
2.8
5
Select Table number
(contrast) and Dmax in
the pairs shown.
Do not use with
Fuji
CR.(See
Fuji
CR below.)
Fuji
CR
FCR302C4.w87
10
11
12
13
14
15
2.6
2.7
2.8
2.9
3.0
3.1
5
Table number
(contrast) and Dmax
must be chosen in the
pairs shown. The
contrast curves are
approximately linear in
density for most of the
pixel value range, but
become relatively flat at
high (bright) pixel
values. Designed for use
with
Fuji
CR only.
CT
(Computed
Tomography) or
MRI
(Magnetic Resonance)
VER693C0.w87
5,
7
, 8
3.0, 3.1
15
Table 7 provides good
grayscale range in all CT
and MRI studies. Table 5
has less mid-tone
contrast. Table 8 has
more mid-tone contrast.
Summary of Contents for DryView 8150
Page 1: ...Kodak DryView 8150 Laser Imager User Guide ...
Page 22: ...Blank Page ...
Page 27: ...October 4 2004 7F3319 5 Dissipating Heat 7 5 Glossary ...
Page 28: ...6 7F3319 October 4 2004 ...
Page 30: ...1 2 7F3319 October 4 2004 Introduction 8150 Laser Imager H200_0036CAA ...
Page 34: ...Blank Page ...
Page 104: ...2 70 7F3319 October 4 2004 Operation and Maintenance Blank Page ...
Page 124: ...Blank Page ...
Page 148: ...4 24 7F3319 October 4 2004 Adding a Modality 8 Touch Yes 9 Wait for the front door to open ...
Page 152: ...Blank Page ...
Page 168: ...Blank Page ...
Page 174: ...Blank Page ...
Page 177: ......